Gen5 System Test Report
Reader: Synergy HT (Serial Number: 194331)
Basecode: P/N 7090202 (v2.25)
Gen5 Version: 2.09.1
Date and Time: 9/8/2022 4:37:55 PM
User: Arnett_Test
Company: Arnett_Supply
Comments: Final Test Run
Test Results
Operator ID:______________________________________________________________
Notes:____________________________________________________________________
__________________________________________________________________________
SYSTEM SELF TEST
7090202 Version 2.25 194331 1100
Bias current offset -0.6 counts PASS
Offset voltage 1539 counts PASS
750V measurement 8.7 counts PASS
750V noise 4 counts
500V measurement 1.0 counts
500V noise 1 counts
Lambda: 200 Gain: 1.84 Resets: 1
Channel: Ref 1
Air: 13738 38783
Dark: 9829 9818
Delta: 3909 28965
Lambda: 352 Gain: 2.61 Resets: 4
Channel: Ref 1
Air: 12849 39853
Dark: 9830 9816
Delta: 3019 30037
Lambda: 620 Gain: 1.87 Resets: 1
Channel: Ref 1
Air: 12560 39968
Dark: 9829 9819
Delta: 2731 30149
Lambda: 790 Gain: 3.01 Resets: 2
Channel: Ref 1
Air: 12544 39850
Dark: 9833 9817
Delta: 2711 30033
Lambda: 860 Gain: 2.12 Resets: 1
Channel: Ref 1
Air: 12463 39085
Dark: 9831 9819
Delta: 2632 29266
Lambda: 962 Gain: 2.56 Resets: 1
Channel: Ref 1
Air: 12518 39869
Dark: 9833 9820
Delta: 2685 30049
Channel: Ref 1
Noise Max: 9868 9812
Noise Min: 9867 9812
Delta: 1 0
Voltage Reference: Lamp 24V Mtr Min Low High Max
1740 1937 2048 1431 1740 2164 2473
INCUBATOR SELF TEST
Temperature Setpoint: 0.0 Current Average: 26.1 A/D Test: PASS
Zone 1: 26.2 Min: 26.1 Max: 26.2 Range: PASS Thermistor: PASS
Zone 2: 26.2 Min: 26.2 Max: 26.2 Range: PASS Thermistor: PASS
Zone 3: 26.0 Min: 26.0 Max: 26.0 Range: PASS Thermistor: PASS
Zone 4: 26.0 Min: 26.0 Max: 26.0 Range: PASS Thermistor: PASS
AUTOCAL ANALYSIS
PROBE: TOP
Upper Left Corner: x= 9696 y= 200
Lower Left Corner: x= 9692 y= 5740
Lower Right Corner: x= 1016 y= 5744
Upper Right Corner: x= 1020 y= 200
Delta 1: 9696 - 9692= +4
Delta 2: 1020 - 1016= +4
Delta 3: 200 - 200= +0
Delta 4: 5744 - 5740= +4
PROBE: BOTTOM
Upper Left Corner: x= 9696 y= 1756
Lower Left Corner: x= 9692 y= 7288
Lower Right Corner: x= 1016 y= 7292
Upper Right Corner: x= 1012 y= 1756
Delta 1: 9696 - 9692= +4
Delta 2: 1012 - 1016= -4
Delta 3: 1756 - 1756= +0
Delta 4: 7292 - 7288= +4
PROBE: ABSORB
Upper Left Corner: x= 11240 y= 1768
Lower Left Corner: x= 11240 y= 7308
Lower Right Corner: x= 2552 y= 7308
Upper Right Corner: x= 2552 y= 1772
Delta 1: 11240 -11240= +0
Delta 2: 2552 - 2552= +0
Delta 3: 1772 - 1768= +4
Delta 4: 7308 - 7308= +0
Probe Height: 34.11
Middle Sensor: y= 11980
Tested: 11984
Delta: +4
Back Sensor: x= 11556 y= 7868
Tested: 11556 7864
Delta: +0 -4
SYSTEM TEST PASS
0000
Excitation Filters: Plug, 485/20, 530/25, 590/20
Emission Filters: 460/40, 528/20, 590/35, 645/40