BioTek
Biotek Agilent Synergy Neo2 Hybrid Multimode Microplate Reader NEO2MB
Biotek Agilent Synergy Neo2 Hybrid Multimode Microplate Reader
Model: Neo2
PN: NEO2MB
Specs:
- MFG: 10/2017
- Monochromator: ABS, Fl, Lum
- Dual filter top: Fl/FP(F Polarization)/TRF and Lum
- Filter bottom: Fl/FP/TRF
Includes:
- NEO2MB Instrument
- AC Adapter
- Win 10 laptop w/ Gen 5 2.09 installed (30 day demo license included, purchaser will need a permanent license from Agilent)
Condition: Used, tested in good working condition. Filter cubes are NOT included. Last serviced 2/22/2022.
WORLDWIDE SHIPPING AVAILABLE
Gen5 System Test Report Reader: Synergy Neo2 (Serial Number: 171009D) Basecode: P/N 1350200 (v1.17) Gen5 Version: 2.09.1 Date and Time: 10/19/2022 4:46:56 PM User: Engineering Company: PNW Scientific Comments: Final Self-Test Test Results SYSTEM TEST PASS Operator ID:______________________________________________________________ Notes:____________________________________________________________________ __________________________________________________________________________ SYSTEM SELF TEST 1350200 Version 1.17 171009D S1 S2 S3 1010 1000 0000 M B Voltage Reference Test Min Low High Max Mono System Flash 1410 1748 2104 3099 0000933888 Top Filter Flash 2340 2923 3506 0000312188 Bottom Filter Flash 2341 2925 3508 0000044324 Switched 24V Power 1880 ABSORBANCE Optics Test Ref Meas Gain Resets R/G #1:230 2.85 2 0.703 Tested 2.94 2 0.679 Light 13478 39849 Dark 10564 10659 Delta 2914 29190 #2:285 2.08 4 1.921 Tested 1.99 4 2.015 Light 13068 39778 Dark 10549 10636 Delta 2519 29142 #3:405 1.55 8 5.154 Tested 1.44 8 5.561 Light 12971 39913 Dark 10541 10622 Delta 2430 29291 #4:630 1.61 8 4.967 Tested 1.57 8 5.092 Light 12661 39866 Dark 10543 10626 Delta 2118 29240 #5:800 1.78 4 2.249 Tested 1.73 4 2.312 Light 12540 39810 Dark 10545 10629 Delta 1995 29181 #6:999 1.84 2 1.085 Tested 1.78 2 1.125 Light 12504 39923 Dark 10545 10631 Delta 1959 29292 Noise Test Ref Meas Gain 1.00 Max 10533 10612 Min 10532 10612 Delta 1 0 FLUORESCENCE/LUMINESCENCE Monochromator System Bias current offset 0.0 counts PASS Offset voltage 1660 counts PASS 750V measurement 167.5 counts PASS 750V noise 85 counts 750V offset 1673 counts Reset offset 1693 counts Bias current 0.00081 nA 1000V current 15.34977 nA Reference bias -0.2 counts PASS Reference offset 10520 counts PASS Reference noise 0.2 counts PASS Top Filter System - Top Bias current offset -0.2 counts PASS Offset voltage 1649 counts PASS 750V measurement 117.4 counts PASS 750V noise 75 counts 750V offset 1656 counts Reset offset 1687 counts Bias current 0.00028 nA 1000V current 5.98956 nA Reference bias -0.3 counts PASS Reference offset 10544 counts PASS Reference noise 0.2 counts PASS Top Filter System - Side Bias current offset 0.0 counts PASS Offset voltage 1653 counts PASS 750V measurement 38.0 counts PASS 750V noise 19 counts 750V offset 1654 counts Reset offset 1694 counts Bias current 0.00021 nA 1000V current 0.53879 nA Bottom Filter System Bias current offset 0.1 counts PASS Offset voltage 1648 counts PASS 750V measurement 415.9 counts PASS 750V noise 183 counts 750V offset 1676 counts Reset offset 1689 counts Bias current 0.00033 nA 1000V current 192.06264 nA Reference bias -0.1 counts PASS Reference offset 10547 counts PASS Reference noise 0.2 counts PASS Filter Fluorescence Top Probe Reference 400V 500V 600V Gain 1.00 1.00 1.00 Light 12072 13188 14523 Dark 10545 10545 10545 Delta 1527 2643 3978 Bottom Probe Reference 400V 500V 600V Gain 1.00 1.00 1.00 Light 12034 13108 14379 Dark 10548 10548 10547 Delta 1486 2560 3832 Mono Fluorescence - Optics Test - 500V Top Probe Bandpass 17nm 40nm Sensitivity:33 Ref Meas Ref Meas #1:300 Light 13426 4189 13981 7977 Dark 10559 1658 10541 1658 Delta 2867 2531 3440 6319 Max 2903 2569 3510 6381 Min 2792 2452 3369 6216 StdDev 30 34 41 52 #2:485 Light 27306 14982 28429 27276 Dark 10559 1658 10541 1655 Delta 16747 13324 17888 25621 Max 16883 13433 18061 25806 Min 16563 13151 17702 25416 StdDev 112 102 110 138 Bottom Probe Bandpass 17nm 40nm Sensitivity:30 Ref Meas Ref Meas #1:300 Light 13462 2673 13999 4332 Dark 10559 1658 10540 1658 Delta 2903 1015 3459 2674 Max 2929 1029 3500 2709 Min 2855 999 3413 2643 StdDev 25 10 28 24 #2:485 Light 27453 21337 28493 36945 Dark 10559 1658 10541 1652 Delta 16894 19679 17952 35293 Max 17041 19797 18201 35671 Min 16798 19618 17516 34846 StdDev 83 56 197 236 CALIBRATION Carrier - Top Mono Fluorescence Upper Left x= -122 y= 8650 Lower Left x= -120 y= 2436 Lower Right x= 9649 y= 2441 Upper Right x= 9646 y= 8655 Delta 1 -122 - -120= -2 Delta 2 9646 - 9649= -3 Delta 3 8655 - 8650= +5 Delta 4 2441 - 2436= +5 Carrier - Bottom Mono Fluorescence Upper Left x= 1859 y=10531 Lower Left x= 1860 y= 4317 Lower Right x=11633 y= 4322 Upper Right x=11628 y=10536 Delta 1 1859 - 1860= -1 Delta 2 11628 -11633= -5 Delta 3 10536 -10531= +5 Delta 4 4322 - 4317= +5 Carrier - Absorbance Upper Left x= 1864 y= 8627 Lower Left x= 1866 y= 2417 Lower Right x=11637 y= 2422 Upper Right x=11635 y= 8633 Delta 1 1864 - 1866= -2 Delta 2 11635 -11637= -2 Delta 3 8633 - 8627= +6 Delta 4 2422 - 2417= +5 Carrier - Top Filter Fluorescence Upper Left x=-3650 y= 6646 Lower Left x=-3647 y= 434 Lower Right x= 6119 y= 441 Upper Right x= 6118 y= 6653 Delta 1 -3650 --3647= -3 Delta 2 6118 - 6119= -1 Delta 3 6653 - 6646= +7 Delta 4 441 - 434= +7 Carrier - Bottom Filter Fluorescence Upper Left x=-3197 y=10057 Lower Left x=-3195 y= 3843 Lower Right x= 6574 y= 3848 Upper Right x= 6571 y=10061 Delta 1 -3197 --3195= -2 Delta 2 6571 - 6574= -3 Delta 3 10061 -10057= +4 Delta 4 3848 - 3843= +5 Carrier - Injectors Upper Left x= 2206 y= 6646 Lower Left x= 2209 y= 434 Lower Right x=11975 y= 441 Upper Right x=11974 y= 6653 Delta 1 2206 - 2209= -3 Delta 2 11974 -11975= -1 Delta 3 6653 - 6646= +7 Delta 4 441 - 434= +7 Carrier - Test Sensors Middle Sensor 23154 Tested 23148 Delta -6 Top Probe Height 25.99 mm Bottom Probe Height 13.90 mm Cube Sliders Top Lower 6388 Top Upper 7220 Bottom 6352 Excitation Monochromator Slit Wheel 8568 Probe Changer 4216 Backlash 20 305LP Edge -969.65 Tested -969.48 Top Fluorescence B=-0.00174951 C=+1.41474628 Bottom Fluorescence B=-0.00152410 C=+1.42230988 Absorbance B=-0.00009621 C=-0.09754116 Emission Monochromator Slit Wheel 8600 Probe Changer 4152 Backlash 24 629nm Peak -52.03 Tested -52.41 Top Fluorescence B=+0.00098055 C=-0.59227067 Bottom Fluorescence B=+0.00178638 C=-1.29126680 Cube Scanners Min Max Top Lower 18 18 Top Upper 20 20 Bottom 20 20 INCUBATION Temperature Setpoint: 0.0 Current Average: 21.0 A/D Test: PASS Zone 1: 21.0 Min: 20.9 Max: 21.0 Range: PASS Thermistor: PASS Zone 2: 20.9 Min: 20.9 Max: 20.9 Range: PASS Thermistor: PASS Zone 3: 21.1 Min: 21.0 Max: 21.1 Range: PASS Thermistor: PASS Zone 4: 21.1 Min: 21.1 Max: 21.1 Range: PASS Thermistor: PASS 0000 Dispenser 1: 005.0,010.0,020.0,040.0,080.0,200.0 Dispenser 2: 005.0,010.0,020.0,040.0,080.0,200.0 Excitation monochromator serial number: 299248 Emission monochromator serial number: 299249 Reviewed/Approved By: _______________________________________ Date: ________________ For Technical Support In the U.S.: BioTek Instruments, Inc. Tel: 800 242 4685 Fax: 802 654 0638 In Europe: BioTek Instruments GmbH Tel: 49(0)7136-9680 Fax: 49(0)7136-968-111 All Others: Tel: 802 655 4040 Fax: 802 654 0638 email: TAC@biotek.com Product support center: http://www.biotek.com/service
Worldwide Shipping
30 Day Guarantee
Secure Checkout