Gen5 System Test Report
Reader: Synergy Neo2 (Serial Number: 171009D)
Basecode: P/N 1350200 (v1.17)
Gen5 Version: 2.09.1
Date and Time: 10/19/2022 4:46:56 PM
User: Engineering
Company: PNW Scientific
Comments: Final Self-Test
Test Results
SYSTEM TEST PASS
Operator ID:______________________________________________________________
Notes:____________________________________________________________________
__________________________________________________________________________
SYSTEM SELF TEST
1350200 Version 1.17 171009D S1 S2 S3
1010 1000 0000
M B
Voltage Reference Test Min Low High Max
Mono System Flash 1410 1748 2104 3099 0000933888
Top Filter Flash 2340 2923 3506 0000312188
Bottom Filter Flash 2341 2925 3508 0000044324
Switched 24V Power 1880
ABSORBANCE
Optics Test Ref Meas Gain Resets R/G
#1:230 2.85 2 0.703
Tested 2.94 2 0.679
Light 13478 39849
Dark 10564 10659
Delta 2914 29190
#2:285 2.08 4 1.921
Tested 1.99 4 2.015
Light 13068 39778
Dark 10549 10636
Delta 2519 29142
#3:405 1.55 8 5.154
Tested 1.44 8 5.561
Light 12971 39913
Dark 10541 10622
Delta 2430 29291
#4:630 1.61 8 4.967
Tested 1.57 8 5.092
Light 12661 39866
Dark 10543 10626
Delta 2118 29240
#5:800 1.78 4 2.249
Tested 1.73 4 2.312
Light 12540 39810
Dark 10545 10629
Delta 1995 29181
#6:999 1.84 2 1.085
Tested 1.78 2 1.125
Light 12504 39923
Dark 10545 10631
Delta 1959 29292
Noise Test Ref Meas
Gain 1.00
Max 10533 10612
Min 10532 10612
Delta 1 0
FLUORESCENCE/LUMINESCENCE
Monochromator System
Bias current offset 0.0 counts PASS
Offset voltage 1660 counts PASS
750V measurement 167.5 counts PASS
750V noise 85 counts
750V offset 1673 counts
Reset offset 1693 counts
Bias current 0.00081 nA
1000V current 15.34977 nA
Reference bias -0.2 counts PASS
Reference offset 10520 counts PASS
Reference noise 0.2 counts PASS
Top Filter System - Top
Bias current offset -0.2 counts PASS
Offset voltage 1649 counts PASS
750V measurement 117.4 counts PASS
750V noise 75 counts
750V offset 1656 counts
Reset offset 1687 counts
Bias current 0.00028 nA
1000V current 5.98956 nA
Reference bias -0.3 counts PASS
Reference offset 10544 counts PASS
Reference noise 0.2 counts PASS
Top Filter System - Side
Bias current offset 0.0 counts PASS
Offset voltage 1653 counts PASS
750V measurement 38.0 counts PASS
750V noise 19 counts
750V offset 1654 counts
Reset offset 1694 counts
Bias current 0.00021 nA
1000V current 0.53879 nA
Bottom Filter System
Bias current offset 0.1 counts PASS
Offset voltage 1648 counts PASS
750V measurement 415.9 counts PASS
750V noise 183 counts
750V offset 1676 counts
Reset offset 1689 counts
Bias current 0.00033 nA
1000V current 192.06264 nA
Reference bias -0.1 counts PASS
Reference offset 10547 counts PASS
Reference noise 0.2 counts PASS
Filter Fluorescence
Top Probe
Reference 400V 500V 600V
Gain 1.00 1.00 1.00
Light 12072 13188 14523
Dark 10545 10545 10545
Delta 1527 2643 3978
Bottom Probe
Reference 400V 500V 600V
Gain 1.00 1.00 1.00
Light 12034 13108 14379
Dark 10548 10548 10547
Delta 1486 2560 3832
Mono Fluorescence - Optics Test - 500V
Top Probe
Bandpass 17nm 40nm
Sensitivity:33 Ref Meas Ref Meas
#1:300
Light 13426 4189 13981 7977
Dark 10559 1658 10541 1658
Delta 2867 2531 3440 6319
Max 2903 2569 3510 6381
Min 2792 2452 3369 6216
StdDev 30 34 41 52
#2:485
Light 27306 14982 28429 27276
Dark 10559 1658 10541 1655
Delta 16747 13324 17888 25621
Max 16883 13433 18061 25806
Min 16563 13151 17702 25416
StdDev 112 102 110 138
Bottom Probe
Bandpass 17nm 40nm
Sensitivity:30 Ref Meas Ref Meas
#1:300
Light 13462 2673 13999 4332
Dark 10559 1658 10540 1658
Delta 2903 1015 3459 2674
Max 2929 1029 3500 2709
Min 2855 999 3413 2643
StdDev 25 10 28 24
#2:485
Light 27453 21337 28493 36945
Dark 10559 1658 10541 1652
Delta 16894 19679 17952 35293
Max 17041 19797 18201 35671
Min 16798 19618 17516 34846
StdDev 83 56 197 236
CALIBRATION
Carrier - Top Mono Fluorescence
Upper Left x= -122 y= 8650
Lower Left x= -120 y= 2436
Lower Right x= 9649 y= 2441
Upper Right x= 9646 y= 8655
Delta 1 -122 - -120= -2
Delta 2 9646 - 9649= -3
Delta 3 8655 - 8650= +5
Delta 4 2441 - 2436= +5
Carrier - Bottom Mono Fluorescence
Upper Left x= 1859 y=10531
Lower Left x= 1860 y= 4317
Lower Right x=11633 y= 4322
Upper Right x=11628 y=10536
Delta 1 1859 - 1860= -1
Delta 2 11628 -11633= -5
Delta 3 10536 -10531= +5
Delta 4 4322 - 4317= +5
Carrier - Absorbance
Upper Left x= 1864 y= 8627
Lower Left x= 1866 y= 2417
Lower Right x=11637 y= 2422
Upper Right x=11635 y= 8633
Delta 1 1864 - 1866= -2
Delta 2 11635 -11637= -2
Delta 3 8633 - 8627= +6
Delta 4 2422 - 2417= +5
Carrier - Top Filter Fluorescence
Upper Left x=-3650 y= 6646
Lower Left x=-3647 y= 434
Lower Right x= 6119 y= 441
Upper Right x= 6118 y= 6653
Delta 1 -3650 --3647= -3
Delta 2 6118 - 6119= -1
Delta 3 6653 - 6646= +7
Delta 4 441 - 434= +7
Carrier - Bottom Filter Fluorescence
Upper Left x=-3197 y=10057
Lower Left x=-3195 y= 3843
Lower Right x= 6574 y= 3848
Upper Right x= 6571 y=10061
Delta 1 -3197 --3195= -2
Delta 2 6571 - 6574= -3
Delta 3 10061 -10057= +4
Delta 4 3848 - 3843= +5
Carrier - Injectors
Upper Left x= 2206 y= 6646
Lower Left x= 2209 y= 434
Lower Right x=11975 y= 441
Upper Right x=11974 y= 6653
Delta 1 2206 - 2209= -3
Delta 2 11974 -11975= -1
Delta 3 6653 - 6646= +7
Delta 4 441 - 434= +7
Carrier - Test Sensors
Middle Sensor 23154
Tested 23148
Delta -6
Top Probe Height 25.99 mm
Bottom Probe Height 13.90 mm
Cube Sliders
Top Lower 6388
Top Upper 7220
Bottom 6352
Excitation Monochromator
Slit Wheel 8568
Probe Changer 4216
Backlash 20
305LP Edge -969.65
Tested -969.48
Top Fluorescence B=-0.00174951 C=+1.41474628
Bottom Fluorescence B=-0.00152410 C=+1.42230988
Absorbance B=-0.00009621 C=-0.09754116
Emission Monochromator
Slit Wheel 8600
Probe Changer 4152
Backlash 24
629nm Peak -52.03
Tested -52.41
Top Fluorescence B=+0.00098055 C=-0.59227067
Bottom Fluorescence B=+0.00178638 C=-1.29126680
Cube Scanners Min Max
Top Lower 18 18
Top Upper 20 20
Bottom 20 20
INCUBATION
Temperature Setpoint: 0.0 Current Average: 21.0 A/D Test: PASS
Zone 1: 21.0 Min: 20.9 Max: 21.0 Range: PASS Thermistor: PASS
Zone 2: 20.9 Min: 20.9 Max: 20.9 Range: PASS Thermistor: PASS
Zone 3: 21.1 Min: 21.0 Max: 21.1 Range: PASS Thermistor: PASS
Zone 4: 21.1 Min: 21.1 Max: 21.1 Range: PASS Thermistor: PASS
0000
Dispenser 1: 005.0,010.0,020.0,040.0,080.0,200.0
Dispenser 2: 005.0,010.0,020.0,040.0,080.0,200.0
Excitation monochromator serial number: 299248
Emission monochromator serial number: 299249
Reviewed/Approved By: _______________________________________ Date: ________________
For Technical Support
In the U.S.:
BioTek Instruments, Inc.
Tel: 800 242 4685
Fax: 802 654 0638
In Europe:
BioTek Instruments GmbH
Tel: 49(0)7136-9680
Fax: 49(0)7136-968-111
All Others:
Tel: 802 655 4040
Fax: 802 654 0638
email: TAC@biotek.com
Product support center: http://www.biotek.com/service
Worldwide Shipping
30 Day Guarantee
Secure Checkout