GE AKTA Explorer 100 FPLC System
GE AKTA Explorer 100 FPLC System
Includes:
- CU-950 control box
- Box-900 drawer w/ tray
- pH/C-900 pH and conductivity monitor
- UV-900 multi-wavelength UV-Vis monitor
- P-900 Pump
Condition: Parts Only, AS-IS
-
"Stack" (Cu-950, BOX-900, pH/C-900, UV-900, P-900)devices all pass self-test.
-
Controller turns on and talks to the system. Controller can talk to a host PC.
-
A/B valves appear to work correctly.
-
Pump pistons and heads appear to seal and work correctly.
-
Performed basic internal and external cleaning of all parts and fittings to remove process materials and any solvents.
Known Issues:
-
Valve 1 has bad motor. Responds to system command but doesn't rotate as it should. Valve rotor and head may also be defective, could not test.
-
Valve 5 has bad motor. Responds to system command but doesn't rotate as it should. Also has damaged valve head- fitting disintegrated in the #1 position and can't be fully removed. Valve rotor may also be defective, could not test.
-
pH sensor does not work correctly.
-
Drain finish area and rear plastic near nameplate are corroded (see pictures for details).
- Other issues may be present.